3.2 Remote control commands

719 S Titrino

93

ON means that the Titrino reports automatically the moment the correspond-

ing change occurs.

.Status:

Global switch for all set AutoInfo.

.P

PowerOn: Simulation of power on (3.2.2.87). Not from mains.

Messages from node .T, Titrator:

.T.R

Ready: Status 'Ready' has been reached.

.T.G

Go: Instrument has been started.

.T.GC

GoCommand: Instrument has received a go command.

.T.S

Stop: Status 'Stop' has been reached.

.T.B

Begin of sequence.

.T.F

Final: End of determination, the final steps will be carried out.

.T.E

Error. Message together with error number, see page 60ff.

.T.H

Hold: Status 'Hold' has been reached.

.T.C

Continue: Continue after hold.

.T.O

Conditioning OK: EP reached (in SET with conditioning).

.T.N

Conditioning Not OK: EP not reached (in SET with conditioning).

.T.Re

Request: In the inquiry of an identification or the sample size after

start of titration.

.T.Si

SiloEmpty: Silo empty, i.e. the last line has been removed from

the silo memory.

.T.EP

EPList: Entry into EP list (with SET)

.T.RC

Results have been recalculated.

Messages for changings in the I/O lines. If the changings are made simulta-

neously, there is 1 message. Pulses receive 2 messages: one message each

for line active and inactive.

.I

Input: Change of an input line.

.O

Output: Change of an output line (except 7, pin 2, for recorder

pulses).

If a change occurs that requires a message, the Titrino sends space (ASCII

32) and ! as an introducer. This is followed by the name of the device (see

3.2.2.47). Special ASCII characters in the device name are ignored. If no de-

vice name has been entered, only ! is sent. Finally the Titrino sends the infor-

mation which node has triggered the message.

Example: !John".T.Si": The message was triggered from instrument "John",

node .T.Si

3.2.2.87. Setup.PowerOn $G

Simulation of 'power on'. The device has the same status as after power on:

The cylinder is filled, error messages deleted and the current sample number

set to 0. The method last used is ready for operation.

3.2.2.88. Setup.Initialise $G

Setup.Initialise.Select ActMeth, Silo, Calib, Config,

Assembly, Setup, All

Setting of default values for the following areas:

ActMeth: Current method. Parameters, calculations, and assignments for

the data output, operands C01...C19.

Silo:

The silo memory is deleted. Same function as delete entire silo.

Calib:

pH calibration data for all measuring inputs.

Config:

All values under &Config.

Assembly: All values under &Assembly.