3.2. Remote control commands

719 S Titrino

76

3.2.2 Description of the remote control commands

3.2.2.1. Mode

$G, $S, $H, $C

Start and stop ($G, $S) or hold of the current method (3.2.2.3) with $H and

continue with $C.

$G also serves to continue after inquiries of identifications and sample size af-

ter the start (see 3.2.2.25) as well as after inquiries of calibration temperature

and pH values of buffers (see 3.2.2.31 and 3.2.2.32).

3.2.2.2. Mode.QuickMeas $G,

$S

Start and stop of a measurement in the basic mode with the parameters

(measured quantity, measuring input) of the current method. Corresponds to

the <meas/hold> key.

With an ongoing measurement, the current mode can be started. This stops

the measurement automatically.

3.2.2.3. Mode.Select SET, MEAS, CAL

Mode.SETQuantity pH, U, Ipol, Upol

Mode.MEASQuantity pH, U, Ipol, Upol, T

Selection of the standard mode. Mode and the measured quantity belong to

the complete selection.

If a method is selected from the method memory, the nodes &Mode.Select

and &Mode.XXXQuantity are overwritten with mode and measured quantity of

the corresponding user method.

3.2.2.4. Mode.Name read

only

Name of the current method in the working memory. $Q sends 8 ASCII char-

acters. Standard methods carry the name ********. The node can be set

read + write, see 3.2.2.61.

3.2.2.5. Mode.Def.Formulas.1.Formula

EPX, CXX, RSX, +, -, *, /, (, )

Mode.Def.Formulas.1.TextRS

up to 8 ASCII characters

Mode.Def.Formulas.1.Decimal 0...2...5

Mode.Def.Formulas.1.Unit

up to 6 ASCII characters

Mode.Def.Formulas.2.Formula

etc. up to .9

Entry of formulas. Rules for formula entry, see page 39f.

Example: "(EP2-EP1)*C01/C00"

In addition to the formula, a text for result output, the number of decimal

places and a unit for the result output can be selected. "No unit" is selected

with the blank string.

In place of "RSX", a result name may be entered (.TextRS). This name is out-

putted in the report full, short. It is used for the result and the corresponding

mean value.

3.2.2.6. Mode.Def.ComVar.C30

RSX, MNX, EPX, CXX

Mode.Def.ComVar.C31

etc., up to .C39