Metrohm 719 S Titrino Manual del usuario

Página 94

Advertising
background image

3.2. Remote control commands

719 S Titrino

90

.Ipol:

Polarization current in µA.

.Upol:

Polarization potential in mV, entry in steps of 10 mV.

3.2.2.76. Assembly.Outputs.AutoEOD ON, OFF

Assembly.Outputs.SetLines $G

Assembly.Outputs.SetLines.L0

active, inactive, pulse, OFF

up to .L 3

Assembly.Outputs.ResetLines $G

Setting the I/O output lines.

.AutoEOD: The automatic output of the EOD (End of Determination) at the

end of the determination can be switched off. Thus, for example,

in conjunction with a Titrino several determinations can be per-

formed in the same beaker. Before AutoEOD is switched on, line 3

must be set to "OFF".

.SetLines: With $G, all lines are set.

.SetLines.LX: Set the line LX. "active" means setting of a static signal, "inac-

tive" means resetting of the signal, "pulse" means output of a

pulse of app. 150 ms, "OFF" means the line is not operated, see

also page 130.

Warnings:
• L2 is the EOD line. If you have "AutoEOD" set to "ON", an active

line 2 is set to "inactive" by the EOD pulse.

• L3 is the line of the activate pulse. An active line 3 is set to "in-

active" by the activate pulse.

.ResetLines: Lines are set to the inactive status (= high).

3.2.2.77. Setup.Keycode ON, OFF

ON means the key code of a key pressed on the Titrino is outputted. The key

code comprises 2 ASCII characters; table of the keys with their code, see

page 108. A keystroke of key 11 is sent as follows:

#11

The beginning of the message is marked by a space (ASCII 32).

3.2.2.78. Setup.Tree.Short ON, OFF

Setup.Tree.ChangedOnly ON, OFF

Definition of the type of answer to $Q.

.Short:

With "ON", each path is sent with only the necessary amount of

characters in order to be unequivocal (printed in bold in this manual). A com-

bination of .Short and .ChangedOnly is not possible.

.ChangedOnly:

Sends only the changed values, i.e. values which have

been edited. All paths are sent absolute, i.e. from the root.

3.2.2.79. Setup.Trace ON, OFF

The Titrino automatically reports when a value has been confirmed with <en-

ter> at the Titrino. Message, e.g.:

&SmplData.OFFSilo.Id1"Trace"

The beginning of the message is marked by a space (ASCII 32).