3.1.3 Status messages

In order to have an efficient control by an external control device, it must also

be possible to query status conditions; they provide information on the status

of the Titrino. The trigger $D initiates output of the status. Status messages

consist of the global status, the detailed status and eventual error messages,

e.g. $S.Mode.SET;E26. The global status informs on the activity of the proc-

ess, while the detailed status conditions show the exact activity within the

process.

The following global status conditions are possible:

$G

Go:

The Titrino is executing the last command.

$H

Hold:

The Titrino has been held ($H, key <meas/hold> or by an error which effects

the hold status)

$C

Continue:

The Titrino has been restarted actively after hold

$R

Ready:

The Titrino has executed the last command and is ready

$S

Stop:

A process has been aborted in an "unnatural manner". e.g. stopped or aborted

because there was an error.

Detailed status conditions

Status conditions of the global $G:

$G .Mode.SET .Inac:

Instrument at the beginning or at the end of a titration.

.Req

.Id1:

Instrument in the SET mode, requesting Id1 after start.

.Id2:

Instrument in the SET mode, requesting Id2 after start.

.Id3:

Instrument in the SET mode, requesting Id3 after start.

.Smpl:

Instrument in the SET mode, requesting sample size after start.

.Unit:

Instrument in the SET mode, requesting unit of sample size after start.

.Start:

Instrument in the SET mode, processing the start conditions.

.SET1:

Instrument in the SET mode, titrating to the first endpoint.

.SET2:

Instrument in the SET mode, titrating to the second endpoint.

.Cond.Ok:

Instrument in the SET, conditioning, endpoint reached (after the first

startup from the standby mode).

.Cond.Prog:

Instrument in the SET mode, conditioning, endpoint not reached

(Conditioning progressing).

$G .Mode.MEAS .Inac:

Instrument at the beginning or at the end of a titration.

.Req .Id1:

Instrument in the MEAS mode, requesting Id1 after start.

.Id2:

Instrument in the MEAS mode, requesting Id2 after start.

.Id3:

Instrument in the MEAS mode, requesting Id3 after start.

.Smpl:

Instrument in the MEAS mode, requesting sample size after start.

.Unit:

Instrument in the MEAS mode, requesting unit of sample size after start.

.Meas:

Instrument in the MEAS mode, measuring.

$G .Mode.CAL .Inac:

Instrument at the beginning or at the end of a calibration

.Req.Temp:

Instrument in the CAL mode, requesting calibration temperature.

.Meas.Temp:

Instrument in the CAL mode, measuring calibration temperature.

.Req.Buf1:

Instrument in the CAL mode, requesting pH of buffer 1.

.Meas.Buf1:

Instrument in the CAL mode, measures buffer 1.

.Req.Buf2:

Instrument in the CAL mode, requesting pH of buffer 2.

.Meas.Buf2:

Instrument in the CAL mode, measures buffer 2.

etc.

$G .Assembly.Bur .Fill:

Buret in filling process

.ModeDis:

Buret in DIS mode

Status conditions of the global $H:

The status message of the action which has been held appears.

If the process is held because a monitored limit has been violated, its status

message is $H.Mode.XXX.XXX.