3.2 Remote control commands

719 S Titrino

77

Assignment of common variables.

The values of the common variables are to be found in &Config.ComVar. They

can be viewed and entered there, see 3.2.2.50.

3.2.2.7. Mode.Def.Report.Assign

SET, MEAS, CAL:

full, short, calc, param, calib, ff

Definition of the report sequence, which is outputted automatically at the end

of the determination. Entries of more than one block have to be separated with

";".

3.2.2.8. Mode.Def.Mean.1.Assign RS1, RSX, EPX, CXX

Mode.Def.Mean.2.Assign

etc., up to .9

Assignment of the statistics calculations. Valid assignments are a requirement

for statistics calculations. In addition, the statistics calculation must be

switched on, see 3.2.2.22. Rules for statistics calculations see page 41.

3.2.2.9. Mode.CFmla

Mode.CFmla.1.Value

0...±999 999

Mode.CFmla.2.Value

etc., up to .19

Calculation constants specific to a method. Stored in the method memory of

the Titrino. Operands specific to the sample (3.2.2.52 and 3.2.2.53) and val-

ues of common variables (3.2.2.50) on the other hand are not stored with the

methods.


3.2.2.10. Mode.Parameter.SET1.EP pH:

0...±20.00, OFF

U,

Ipol:

0...±2000, OFF

Upol:

0...±200.0, OFF

Mode.Parameter.SET1.UnitEp read

only

Parameters for SET: Setting the 1st endpoint as pH value, in mV (with U and

Ipol) resp. µA (with Upol). The corresponding unit can be read with .UnitEP.

If the value is on "OFF", no further nodes will appear from SET1.

3.2.2.11. Mode.Parameter.SET1.Dyn pH:

0.01...20.00, OFF

U,

Ipol:

1...2000, OFF

Upol:

0.1...200.0, OFF

Mode.Parameter.SET1.UnitDyn read

only

Mode.Parameter.SET1.MaxRate 0.01...10...150, max.

Mode.Parameter.SET1.MinRate 0.01...25.0...9999.9

Parameters for SET: Control parameters, see page 25.

.Dyn:

Dynamics, control range in pH, mV (with U and Ipol) or µA (with

Upol). The corresponding unit can be read with .UnitDyn.

.MaxRate: Maximum allowed titration rate in mL/min. Max. means maximum

possible rate with the Exchange Unit in current use.

.MinRate: Minimum titration rate in ul/min.