3.1. General rules

719 S Titrino

62

E120

Overrange of the primary measured value (pH, U, Ipol, Upol or T

with MEAS T). The secondary measured value (temperature) may

be instable as well.

Exit: Correct error or &m $S.

E123

Missing EP for calculation.

Exit: The error message disappears on next startup or on

recalculation.

E128

No new mean.

Exit: The error message disappears on next startup or on

recalculation.

E129

No new common variable, old value remains.

Exit: The error message disappears on next startup or on

recalculation.

E130

Wrong sample. For SET with preset titration direction the first

measured value lies behind the endpoint.

Exit: The error message disappears on next startup.

E131

No EP set for SET.

Exit: The error message disappears on next startup.

E132

Silo empty and it has been started with open silo or empty silo has

been opened.

Exit: Send a silo entry.

E133

Silo full.

Exit: Send new command.

E134

No method. A method, which is required from the silo memory, does not

exist.

Exit: The error message disappears on next startup.

E135

Check temp.sensor in MEAS T.

Exit: Correct error or &m $S.

E136

Same buffer in CAL. Measured value of the second buffer differs less than 6

mV from the measured value of the first buffer.

Exit: Correct error or &m $S.

E137

XXX Bytes are missing so that the method, the silo line could not be stored.

Exit: Send new command.