Advertising
background image

3.2. Remote control commands

719 S Titrino

84

$G sets all RS settings. The changes are performed only if the instrument is

inactive. After the setting of the interface parameters, wait at least 2 s to allow

the components to equilibrate.

Settings of the values for the data transmission via the RS interface: baud

rate, data bit, stop bit, parity and type of handshake, see also page 95ff.

The setting of the values must be initiated with $G immediately after entry of

the values.

3.2.2.50. Config.ComVar.C30

with up to .C39, etc.

0... +999 999

Values of the common variables from C30 up to C39. Insert the common

variables directly or describe the determination results directly from the

method, see 3.2.2.6.

3.2.2.51. SmplData.Status ON, OFF

On/off switching of silo memory. When the silo memory is switched on, the

sample data are fetched from the lowest valid silo line.

3.2.2.52. SmplData.OFFSilo.Id1

up to 8 ASCII characters

SmplData.OFFSilo.Id2

up to 8 ASCII characters

SmplData.OFFSilo.Id3

up to 8 ASCII characters

SmplData.OFFSilo.ValSmpl

6-digits, sign and decimal point

SmplData.OFFSilo.UnitSmpl

up to 5 ASCII characters

Current sample data.

The identifications Id1...Id3 can be used in formulas as sample-specific calcu-

lation constants C21...C23.

If "no unit" is desired for the unit of the sample size, the blank string must be

entered.

3.2.2.53. SmplData.ONSilo.Counter.MaxLines read

only

SmplData.ONSilo.Counter.FirstLine read

only

SmplData.ONSilo.Counter.LastLine read

only

Information on silo memory.

.MaxLines: Maximum possible number of silo lines.

.FirstLine: Lowest valid silo line.

.LastLine: Last occupied silo line.

3.2.2.54. SmplData.ONSilo.EditLine.1.Method

up to 8 ASCII characters

SmplData.ONSilo.EditLine.1.Id1

up to 8 ASCII characters

SmplData.ONSilo.EditLine.1.Id2

up to 8 ASCII characters

SmplData.ONSilo.EditLine.1.Id3

up to 8 ASCII characters

SmplData.ONSilo.EditLine.1.ValSmpl 6-digits, sign and dec.point

SmplData.ONSilo.EditLine.1.UnitSmpl

up to 5 ASCII characters

etc., up to .99

Contents of a silo line.

.Method: Method used to process the sample, from the method memory or

from the card.

.Id:

The identifications Id1...Id3 can also be used as sample-specific

calculation constants C21...C23 in formulas.