2 triggers – Metrohm 719 S Titrino Manual del usuario

Página 61

Advertising
background image

3.1 General rules

719 S Titrino

57

3.1.2 Triggers

Triggers initiate an action on the Titrino, for example, starting a process or

sending data. Triggers are marked by the introductory symbol $.

The following triggers are possible:

$G

Go

Starts processes, for ex. starting the mode run or setting the RS 232 interface

parameters

$S Stop Stops

processes

$H Hold Holds

processes

$C Continue

Continues processes after Hold

$Q Query

Queries all information from the current node in the tree forward up to and

including the values

$Q.P Path

Queries the path from the root of the tree up to the current node

$Q.H Highest

Queries the number of son nodes of the current node

Index

$Q.N"i" Name

Queries the name of the son node with index i, i = 1 - n

$D Detail-Info Queries the detailed status information

$U

qUit

Aborts the data flow of the instrument, for example, after $Q

The triggers $G and $S are linked to particular objects, see the summary table

page 63ff.

All other triggers can be used at any time and at all locations on the object

tree.

Examples:

Querying the value of the baud rate: &Config.RSSet.Baud $Q

Querying all values of the node "RSSet": &Config.RSSet $Q

Querying the path of the node "RSSet": &Config.RSSet $Q.P

Start mode: &Mode $G

Querying the detailed status: $D