Metrohm 719 S Titrino Manual del usuario

Página 65

Advertising
background image

3.1 General rules

719 S Titrino

61

RS receive errors:

E36 Parity

Exit: <QUIT> and ensure settings of appropriate parameters at

both devices are the same.

E37 Stop

Bit

Exit: <QUIT> and ensure settings of appropriate parameters at

both devices are the same.

E38

Overrun error. At least 1 character could not be read.

Exit: <QUIT>

E39

The internal working-off buffer of the Titrino is full (>82

characters).

Exit: <QUIT>



RS send errors:

E40

DSR=OFF No proper handshake for more than 1 s.

Exit: <QUIT> Is the receiver switched on and ready to receive?

E41

DCD=ON No proper handshake for more than 1 s.

Exit: <QUIT> Is the receiver switched on and ready to receive?

E42

CTS=OFF No proper handshake for more than 1 s.

Exit: <QUIT> Is the receiver switched on and ready to receive?

E43

The transmission of the Titrino has been interrupted with XOFF for

at least 6 s.

Exit: Send XON or <QUIT>

E44

The RS interface parameters are no longer the same for both

devices. Reset.

E45

The receive buffer of the Titrino contains an incomplete command

(L

F

missing). Sending from the Titrino is therefore blocked.

Exit: Send L

F

or <QUIT>.

Advertising