3.2. Remote control commands

719 S Titrino

86

Info.CalibrationData.Inp1.Date read

only

Info.CalibrationData.Inp1.ElectrodeId read

only

identical for .Inp2 and .Diff

pH calibration data for measuring input 1. After the calibration, the data are

entered automatically together with the date of the calibration and the elec-

trode identification, see 3.2.2.34.

Calibration data can be entered. They are accepted with &Info.CalibrationData

$G. If calibration data are entered, the calibration date is deleted.

3.2.2.60. Info.Checksums $G

Info.Checksums.ActualMethod read

only

The checksums can be used to identify the content of a file unequivocally,

e.g. files with identical content

have identical results of the checksums. An empty file has checksum "0". The

calculation of the checksums is triggered with $G.

.ActualMethod: Result of the checksum of the current method in the working

memory. Identical methods with different method names have the

same results of the checksum.

3.2.2.61. Info.DetermData $G

Info.DetermData.Write ON,

OFF

Info.DetermData.ExV read

only/read+write

Recalculation of the measuring data is triggered with $G.

.Write:

With "ON", the following nodes can be overwritten:

&Info.TitrResults.Var.C4X (X = 0...5) and &Mode.Name.

.ExV:

Volume of the exchange unit, with which the determination was

executed

3.2.2.62. Info.TitrResults.RS.1.Value read

only

etc., up to .9

Info.TitrResults.EP.1.V read

only

Info.TitrResults.EP.1.Meas read

only

etc., up to .2

Info.TitrResults.Var.C40 read

only/read+write

etc., up to .C47

.RS: Values of the calculated results.

.EP: Endpoints with SET:

Volume coordinate in mL, e.g. "1.2340"

Measured value coordinate in pH "5.12", mV (with U and Ipol) "-

241" or µA (with Upol) "43.7".

.Var: Various variables. You may overwrite the variables C40...C45, see

3.2.2.95.

C40: Initial measured value in pH "5.12", mV (with U and Ipol)

"41", µA (with Upol) "43.7" or °C (with T) "25.0". In MEAS final

measured value.

C41: End volume with SET in ml, "12.5360".

C42: Time from start of titration to end in s, "62".

C43: Volume drift on start of a SET titration from the conditioning

in ul/min, "3.5".