3.2. Remote control commands

719 S Titrino

80

delete all: Clear entire statistics table. The results can not be reac-

tivated.

.ResTab.DelN: Specification of the line number to be deleted.

3.2.2.23. Mode.Parameter.Presel.Cond ON, OFF

Mode.Parameter.Presel.DriftDisp ON, OFF

Parameters for SET:

.Cond: Conditioning ON/OFF

.DriftDisp: Drift display during conditioning ON/OFF.

3.2.2.24. Mode.Parameter.Presel.DCor.Type auto,

man, OFF

Mode.Parameter.Presel. DCor.Value 0.0, 99.9

Parameters for SET:

.Type:

Mode of drift correction

.Value:

drift value in µL/min.

If "auto" is set, the actual drift value at the start of titration is used for drift cor-

rection.

Formula for drift correction: EP – (drift value * titration time)

3.2.2.25. Mode.Parameter.Presel.IReq

id1, id1&2, all, OFF

Mode.Parameter.Presel.SReq

value, unit, all, OFF

Parameters for SET, MEAS: Automatic inquiry after the start of the determina-

tion. From such an inquiry, the determination continues if the requested en-

try/entries is/are made, e.g. &SmplData.OFFSilo.Id1 (see 3.2.2.73) or with &M

$G, see 3.2.2.1.

$H is not possible in requests.

3.2.2.26. Mode.Parameter.Presel.ActPuls ON, OFF

Parameter for SET:

first, all, cond., OFF

Output of a pulse on the I/O line "Activate", see page 131.

3.2.2.27. Mode.Parameter.Measuring.SignalDrift

pH, U, Ipol, T:

0.5...999, OFF

Upol: 0.05...99.9, OFF

Mode.Parameter.Measuring.UnitSigDrift read

only

Mode.Parameter.Measuring.EquTime 0...9999, OFF

Parameters for MEAS: Criteria for the measured value acquisition. Measured

value drift in mV/min (with pH, U, Ipol), µA/min (with Upol), resp. °C/min

(with T). Equilibration time in s. OFF means that the corresponding criterion is

switched off. If both criteria are OFF, the measurement continues indefinitely.

If the equilibration time has never been edited, it is automatically calculated by

the instrument to match the drift, see page 34. After it has been edited once, it

remains in force with the set value.

3.2.2.28. Mode.Parameter.Measuring.MeasInput

1, 2, diff.

Mode.Parameter.Measuring.Ipol ±127...1...+127

Mode.Parameter.Measuring.Upol ±1270...400...+1270