Metrohm 719 S Titrino Manual del usuario

Página 95

Advertising
background image

3.2 Remote control commands

719 S Titrino

91

3.2.2.80. Setup.Lock.Keyboard ON, OFF

Setup.Lock.Config ON, OFF

Setup.Lock.Parameter ON, OFF

Setup.Lock.SmplData ON, OFF

Setup.Lock.UserMeth.Recall ON, OFF

Setup.Lock.UserMeth.Store ON, OFF

Setup.Lock.UserMeth.Delete ON, OFF

Setup.Lock.Display ON, OFF

ON means disable the corresponding function:

.Keyboard: Disable all keys of the Titrino

.Config:

Disable the <configuration> key

.Parameter: Disable the <parameter> key

.SmplData: Disable the <smpl data> key

.UserMeth.Recall: Disable "recall" in <user meth> key

.UserMeth.Store: Disable "store" in <user meth> key

.UserMeth.Delete: Disable "delete" in <user meth> key

.Display: Disable the display, i.e. it will not be written to by the device pro-

gram of the Titrino and can be operated from the computer.

3.2.2.81. Setup.Mode.StartWait ON, OFF

Setup.Mode.FinWait ON, OFF

Holding points in the method sequence. If they are "ON", the sequence stops

until "OFF" is sent. Switching the instrument on sets both nodes to OFF:

.StartWait: Holding point right after starting a method or submethod in TIP

(holding point after AutoInfo !".T.GC").

.FinWait: Holding point at the end a method or submethod in TIP (holding

point after AutoInfo !".T.F").

3.2.2.82. Setup.SendMeas.SendStatus ON, OFF

Setup.SendMeas.Interval 0.08...4...16200, MPList

.SendStatus: ON means the automatic transmission of measured values (see

3.2.2.84 and 3.2.2.85) in the inputted interval is active.

.Interval: Time interval (in s) for the automatic transmission of associated

measured values defined under points 3.2.2.84 and 3.2.2.85. The

inputted value is rounded off to a multiple of 0.08. The smallest

possible time interval depends on the number of measured values

which have to be sent, on the baud rate, on the load on the inter-

face and on the type of device connection.

The automatic transmission is switched on/off with 'SendStatus'.

3.2.2.83. Setup.SendMeas.Select Assembly,

Titrator

Selection of the unit of which the measured values should be sent (3.2.2.84

or 3.2.2.85).

3.2.2.84. Setup.SendMeas.Assembly.CyclNo ON, OFF

Setup.SendMeas.Assembly.V ON, OFF

Setup.SendMeas.Assembly.Meas ON, OFF