3.2. Remote control commands

719 S Titrino

82

Parameters for CAL: Calibration at Titrino.

With "ON", there are no hold points in the calibration sequence for entries, the

first buffer is measured directly.

3.2.2.36. Mode.Parameter.Calibration.ActPulse first,

all, OFF

Parameters for CAL: Output of a pulse on the I/O line "Activate", see page

131.

3.2.2.37. UserMeth.FreeMem read

only

Memory space, available for user methods or silo lines. $Q sends the number

of free bytes, e.g. "4928".

3.2.2.38. UserMeth.Recall $G

UserMeth.Recall.Name

up to 8 ASCII characters

UserMeth.Store $G

UserMeth.Store.Name

up to 8 ASCII characters

UserMeth.Delete $G

UserMeth.Delete.Name

up to 8 ASCII characters

UserMeth.DelAll $G

Management of the internal method memory: Load, store and delete methods.

An action is performed if "$G" is sent to the corresponding node just after en-

tering the name.

Do not use blank characters before and after method name!

.DelAll: Deletes all methods in the user memory.

3.2.2.39. UserMeth.List.1.Name read

only

UserMeth.List.1.Mode read

only

UserMeth.List.1.Quantity read

only

UserMeth.List.1.Bytes read

only

UserMeth.List.1.Checksum read

only

for each method

List of the methods in the user method memory with the following character-

istics:

.Name:

Name of the method

.Mode: Mode

.Quantity: Measured

quantity

.Bytes:

Number of bytes of the user memory used by the method

.Checksum: Checksum of the method, see 3.2.2.59.

3.2.2.40. Config.PeriphUnit.CharSet

Epson, Seiko, Citizen, HP, IBM

Selection of the character set and the graphics control characters of the Ti-

trino.

IBM means the IBM character set following character set table 437 and IBM

graphics control characters. Select 'IBM' for work with the computer.

3.2.2.41. Config.PeriphUnit.Balance Sartorius,Mettler,Mettler AT,

AND,Precisa

Config.PeriphUnit.Plot

U, dU/dt, V, dV/dt, U(rel), T