3.1. General rules

719 S Titrino

60

3.1.4 Error messages

Error messages are added to the status messages and separated from them

by the sign ";".

E20

Check exchange unit.

Exit: Mount Exchange Unit (properly) or &m $S.

E21

Check electrode, short circuit.

Exit: Rectify fault or &m $S.

E22

Check electrode, break.

Exit: Rectify fault or &m $S.

E23

Division by zero.

Exit: The error message disappears on next startup or on recalculation.

E26

Manual stop.

Exit: The error message disappears on next startup.

E27

Stop V reached in SET.

Exit: The error message disappears on next startup.

E28

Wrong object call up

Exit: Send correct path for object. Start path at root.

E29

Wrong value or no value allowed.

Exit: Send correct value or call up new object.

E30

Wrong trigger, this trigger is not allowed or carrying-out of action not

possible.

Exit: Send correct trigger (exception: $D) or call up new object.

E31

Command is not possible in active status. Repeat command in inactive

status.

Exit: Send new command.

E32

Command is not possible during titration. Repeat command during the

conditioning phase or in inactive status.

Exit: Send new command.

E33

Value has been corrected automatically.

Exit: Send new command.

E34

Instrument at the end of the titration and sample data is edited; the instrument

at rest or editing during filling.

Exit: &m $S.