719 S Titrino

78

3.2.2.12. Mode.Parameter.SET1.Stop.Type drift, time

Mode.Parameter.SET1.Stop.Drift 1...20...999

Mode.Parameter.SET1.Stop.Time 0...10...999, inf

Mode.Parameter.SET1.Stop.StopT 0...99

999, OFF

Parameters for SET: Type and size of the stop criterion of the titration.

.Type: Type of stop criterion after stop drift or switch-off delay time.

.Drift: Stop drift in µL/min. Applies when "drift" has been selected.

.Time: Switch-off delay time in s. Applies when "time" has been selected. "inf"

means infinite.

.StopT: Stop time in s. Applies when "time" has been selected and the value of

.Time is set to "inf.".

3.2.2.13. Mode.Parameter.TitrPara.Direction +,

-, auto

Parameters for SET: Titration direction.

"auto" means the titration direction is determined automatically by the instru-

ment. If 2 EP's have been set in a SET titration, the titration direction is given

by the two EP's. The entry of the titration direction is then invalid.

3.2.2.14. Mode.Parameter.TitrPara.XPause

0...999 999

Parameter for SET: Pause time 1 in s. Is waited off before the dispensing of

the start volume.

3.2.2.15. Mode.Parameter.TitrPara.StartV.Type abs.,

rel.,

OFF

Mode.Parameter.TitrPara.StartV.V 0...999.99

Mode.Parameter.TitrPara.StartV.Factor

0...±999 999

Mode.Parameter.TitrPara.StartV.Rate 0.01...150, max.

Parameters for SET: Start volume.

If an absolute start volume (abs.) has been selected, the volume in mL is

valid.

A relative start volume (rel.) is dispensed as a function of the sample size:
Start volume in mL = smpl size

*

factor

The factor is valid.

The dispensing rate in mL/min applies to both cases. Max. means maximum

possible dispensing rate with the Exchange Unit in current use.

3.2.2.16. Mode.Parameter.TitrPara.Pause

0...999 999

Parameter for SET: Pause time 2 in s. Is waited off after the dispensing of the

start volume.

3.2.2.17. Mode.Parameter.TitrPara.ExtrT

0...999 999

Parameter for SET: Extraction time in s.

3.2.2.18. Mode.Parameter.TitrPara.MeasInput

1, 2, diff.

Mode.Parameter.TitrPara.Ipol -127...1...+127

Mode.Parameter.TitrPara.Upol -1270...400...+1270

Mode.Parameter.TitrPara.PolElectrTest ON, OFF

Parameters for SET: