3.2 Remote control commands

719 S Titrino

81

Mode.Parameter.Measuring.PolElectrTest ON, OFF

Parameters for MEAS:

Selection of the measuring input; valid with measured quantities pH and U.

"diff." means differential amplifier, see page 124.

With Ipol, the inquiries for the polarization current in µA (Ipol) and

.PolElectrTest are valid.

With Upol the inquiry for the polarization voltage in mV (Upol) is valid. Entry in

steps on 10 mV.

Besides .PolElectrTest is valid.

If the test for polarized electrodes is switched on, it is performed on change-

over from the inactive state to the measurement.

3.2.2.29. Mode.Parameter.Measuring.Temp -170.0...25.0...500.0

Parameters for MEAS: Measurement temperature in °C. If a Pt100 or Pt1000

is connected, the temperature is measured.

The temperature is used for the temperature correction in pH measurements.

3.2.2.30. Mode.Parameter.Calibration.MeasInput

1, 2, diff.

Parameters for CAL: Selection of the measuring input. "diff." means differen-

tial amplifier, see page 124.

3.2.2.31. Mode.Parameter.Calibration.CalTemp -20.0...25.0...120.0

Parameters for CAL: Calibration temperature in °C. If a Pt 100 or Pt1000 is

connected, the temperature is measured.


3.2.2.32. Mode.Parameter.Calibration.Buffer.1.Value 0...7.00...±20.00

Mode.Parameter.Calibration.Buffer.2.Value 0...4.00...±20.00, OFF

etc. up to 9 buffers

Parameters for CAL: pH of buffers. The first buffer which is set to "OFF" de-

termines the number of buffers in the calibration.

3.2.2.33. Mode.Parameter.Calibration.SignalDrift 0.5...2...999, OFF

Mode.Parameter.Calibration.EquTime 0...110...9999, OFF

Parameters for CAL: Criteria for measured value acquisition. Measured value

drift in mV/min, equilibration time in s. OFF means that the corresponding cri-

terion is switched off. If both criterions are on OFF, the measured value is ac-

quired immediately.

If the equilibration time has never been edited, it is automatically calculated by

the instrument to match the drift, see page 34. After it has been edited once, it

remains in force with the set value.

3.2.2.34. Mode.Parameter.Calibration.ElectrodeId

up to 8 ASCII char

Parameters for CAL: Electrode identification. It is classified under calibration

data, see 3.2.2.59.

3.2.2.35. Mode.Parameter.Calibration.SmplChanger ON, OFF