Smpldata – Metrohm 719 S Titrino Manual del usuario

Página 73

Advertising
background image

3.2 Remote control commands

719 S Titrino

69

&SmplData

Object Description

Input

range

Reference

& Root

:

SmplData

Sample data

:

.Status

Status of silo memory

ON, OFF 3.2.2.51.

.OFFSilo

Current sample data

│ ├

.Id1

Sample identification 1

up to 8 ASCII char

3.2.2.52.

│ ├

.Id2

Sample identification 2

up to 8 ASCII char

ditto

│ ├

.Id3

Sample identification 3

up to 8 ASCII char

ditto

│ ├

.ValSmpl Sample

size

±X.XXXXX

ditto

│ ├

.UnitSmpl

Unit of sample size

up to 5 ASCII char

ditto

.ONSilo

Current sample data

│ ├

.Counter

Counter of silo memory

│ │ ├

.MaxLines

Maximum lines

read only

3.2.2.53.

│ │ ├

.FirstLine

First line

read only

ditto

│ │ ├

.LastLine

Last line

read only

ditto

│ ├

.EditLine

Editing silo lines

│ │ ├

.1

1

st

silo line

│ │ │ ├

.Method

Method name

up to 8 ASCII char

3.2.2.54.

│ │ │ ├

.Id1

Sample identification 1

up to 8 ASCII char

ditto

│ │ │ ├

.Id2

Sample identification 2

up to 8 ASCII char

ditto

│ │ │ ├

.Id3

Sample identification 3

up to 8 ASCII char

ditto

│ │ │ ├

.ValSmpl Sample

size

±X.XXXXX

ditto

│ │ │ ├

.UnitSmpl Unit of sample size

up to 5 ASCII char

ditto

│ │ ├

up to 99 lines

│ ├

.DelLine

Delete silo line

$G

3.2.2.55.

│ │ ├

.LineNum Line

number

1...99,

OFF

ditto

│ ├

.DelAll

Delete silo line

$G

3.2.2.56.

│ ├

.CycleLines Cycle

lines

ON,

OFF 3.2.2.57.

Advertising