3.2. Remote control commands

719 S Titrino

88

3.2.2.66. Info.ActualInfo.Assembly.Counter.V read

only

Info.ActualInfo.Assembly.Counter.Clear $G

$Q sends the volume. With the function &Info.Assembly.Counter.Clear $G, the

volume counter is set to zero.

3.2.2.67. Info.ActualInfo.Assembly.Meas read

only

Info.ActualInfo.Assembly.PistonPos read

only

$Q sends the current measured value from the assembly.

Piston position:

0

=

initial position (filled)

10 000 =

end position (empty)

3.2.2.68. Info.ActualInfo.Titrator.CyclNo read

only

Info.ActualInfo.Titrator.V read

only

Info.ActualInfo.Titrator.Meas read

only

Info.ActualInfo.Titrator.dVdt read

only

Info.ActualInfo.Titrator.dMeasdt read

only

Info.ActualInfo.Titrator.dMeasdV read

only

$Q sends the current values in the following formats:

SET

MEAS

CAL

CyclNo 127 127 127
V(ml) 1.2345

- -

Meas:

pH

U, Ipol (mV)

Upol (uA)

T (°C)

3.6(mV)

-345.6

-12.5

-

3.345

-345.6

-12.5

25.0

3.345

-

-

-

dVdt (ul/s)

2.5142

-

-

dMeasdt

pH,U,Ipol mV/s

Upol (uA/s)

T (°C/s)

0.7957

0.7957

-

0.7957

0.7957

0.7957

0.7957

-

-

dMeasdV (mV/ul) 10.6326

-

-

NV: Not Valid. If in the signal drift is OFF in modes MEAS and CAL, the signal

drift is NV.

OV will be sent for "overrange".

A time frame can be set up from the cycle number and the cycle time (see

3.2.2.71). The cycle number is set to 0 at the start of a method and it is in-

cremented until the end of the method.

3.2.2.69. Info.ActualInfo.EP.Index read

only

Info.ActualInfo.EP.X read

only

Info.ActualInfo.EP.Y read

only

.EP.X"1.234" Volume coordinate of the EP

.EP.Y"5.34" Measured value coordinate of the EP

3.2.2.70. Info.ActualInfo.Display.L1

up to 24 ASCII characters