5.12.4 What to do if Data Transfer fails?

Problem

Helpful Questions for Troubleshooting

No characters can be received
on a connected printer.

− Are the instruments switched on and the connection

cables plugged in properly?

− Is the printer set to "on-line"?

− Are baud rate, data bit and parity set identically on both

instruments?

− Is the handshake set properly?

No data transfer occurs and an
error message appears in the
display.

E40–42: Transmission error. Is the cable used correctly

wired and plugged in? Is the printer switched on and set
to "on-line"?

E43: Data output disabled for more than 3 s by XOFF.

E36–39: Receive error. Are the RS232 parameters set

identically on both instruments?

The characters received are
garbled.

− Are data bit and parity set identically for both instruments?

− Is the baud rate set the same on both devices?

− Is the correct printer selected?

− The data transmission was interrupted (by the hardware)

during printing. Re-establish connections again, turn
printer off and on.