5.11 Description of the remote control commands

.Keyboard

Disables all keys of the keyboard, except the
<START>, STOP> and <CLEAR> key.

.Config

Locks the configuration menu

.Parameter

Locks the parameter menu

.Usermeth.Recall

Locks the function "recall method"

.UserMeth.Store

Locks the function "store method"

.Usermeth.Delete

Locks the function "delete method"

.Display

Disables the LCD display. The instrument will not sup-
port the display.

5.11.4.6

Setup.Mode.StartWait on,

off

Indefinite start delay. Only for remote control.

5.11.4.7

Setup.Mode.SendMeas.Sendstatus on,

off

Setup.Mode.SendMeas.Intervall 1…4…999
Automatic sending of the current measured values defined below at a se-
lectable interval (in seconds).

5.11.4.8

Setup.Mode.SendMeas.Meas.CyclNo on,

off

Setup.Mode.SendMeas.Meas.OvenTemp on,

off

Setup.Mode.SendMeas.Meas.GasFlow on,

off

Definition of the measured values to be sent automatically, see above.

5.11.4.9

Setup.AutoInfo.Status

on, off

Setup.AutoInfo.P on,

off

Setup.AutoInfo.Ch.G

on, off

Setup.AutoInfo.Ch.GC on,

off

Setup.AutoInfo.Ch.R on,

off

Setup.AutoInfo.Ch.S

on, off

Setup.AutoInfo.Ch.H

on, off

Setup.AutoInfo.Ch.C

on, off

Setup.AutoInfo.Ch.B

on, off

Setup.AutoInfo.Ch.F on,

off

Setup.AutoInfo.Ch.OM on,

off

Setup.AutoInfo.Ch.CM

on, off

Setup.AutoInfo.Heater.T

on, off

Setup.AutoInfo.Heater.S

on, off

Setup.AutoInfo.E on,

off

"on" means, the instrument automatically sends a corresponding message
when the specified event occurs.
.Status

Enables/disables the preset AutoInfo messages

.P

PowerOn: Simulation of PowerOn (see 5.11.4.10). Not from
mains.

Messages from changer functions:
.Ch.G

Go: Method has been started

.Ch.GC

Go Command: Start command has been received

.Ch.R

Ready: Status 'Ready' reached

.Ch.S

Stop: Status 'Stop' reached

.Ch.H

Hold: Status 'Hold' reached

.Ch.C

Continue: Resuming after Hold

.Ch.B

Begin: Begin of sample sequence

.Ch.F

Final: End of sample sequence

.Ch.OM

"Opening Moves": Begin of start sequence

.Ch.CM

"Closing Moves": Begin of final sequence

.Heater.T Target:

Change

of target temperatur