5.11 Description of the remote control commands

5.11.2.5

Config.RackDef.SpezBeak.1.Pos

0...number of rack positions

etc.

until

.8

Rack positions of special beakers 1 to 8 (position 0 = not defined).

5.11.2.6

Config.PosTab.TabIdx

0...31

Index of position tables..

5.11.2.7

Config.PosTab.Name

8 ASCII characters

Config.PosTab.R1Num

2...(R2Num – 2)

Config.PosTab.R2Num

(R1Num + 2)...(R3Num – 2)

Config.PosTab.R3Num

(R2Num + 2)...200

Config.PosTab.R1Off

0...3599

Config.PosTab.R2Off

0...3599

Config.PosTab.Num

1...200

Definitions of position tables. Depending on the table index the correspond-
ing one of the 16 possible data sets is overlayed.
Name:

Identification of the rack type. Will be available as a selec-
tor under &Config.RackDef.Type (see 5.11.2.4).

R1Num:

Highest Beaker position in row 1

R2Num:

Highest Beaker position in row 2

R3Num:

Highest Beaker position in row 3

R1Off Offset

in

1

/

10

-angular degrees for the beaker positions in

row 1 (for the beaker test)

R2Off Offset

in

1

/

10

-angular degrees for the beaker positions in

row 2 (for the beaker test)

Num:

Number of rack positions

5.11.2.8

Config.PosTab.1.Value

0...3599

bis

.200.Value

Angular offset for the particular rack position in tenth of angular degrees
(

1

/

10

-degrees).

Definition of new rack types:


The definition of rack types is only possible via RS232 interface.

Set table index (see 5.11.2.6).

• Enter name of rack type (&Config.PosTab.Name, see 5.11.2.7).

• Enter highest beaker position in row 1 to 3 (&Config.PosTab.R1–3Num,

see 5.11.2.7).

• Enter the offset angle between beaker position 1 in row 1 and beaker

position (R1num+1) in row 2, respectively and tower 1
(&Config.PosTab.R1–2Off, see 5.11.2.7).

• Define number of rack positions (&Config.Pos.Tab.Num, see 5.11.2.7).

• Enter the particular angular offset for each rack position

(Irregular layouts of rack positions are possible)

5.11.2.9

Config.WetPart.WetPartNo

1...12

Identification of dosing unit.

5.11.2.10

Config.WetPart.1.MaxRate 0.01...160 ml/min

Config.WetPart.1.Length 0...1000...30000 mm

Config.WetPart.1.Diameter 0.1...2...20 mm

until

.4.Diameter

(Default value .2.Length : 250 mm)