Metrohm 774 Oven Sample Processor Manual del usuario

Página 108

Advertising
background image

5 Descripción detallada

RS Send Errors:

E40

DSR=OFF. No proper handshake for more than 1s. Exit: <QUIT>. Is the
receiver switched on and ready to receive?

E41

DCD=ON. No proper handshake for more than 1s. Exit: <QUIT>. Is the
receiver switched on and ready to receive?

E42

CTS=OFF. No proper handshake for more than 1s. Exit: <QUIT>. Is the
receiver switched on and ready to receive?

E43

Transmission has been interrupted for at least 3 s with XOFF.
Exit: send XON or <QUIT>.

E44

The RS-interface parameters are no longer identical for both instruments.
Re-set.

E45

The receive buffer of the instrument contains an incomplete command (LF
missing). Transmission is therefore blocked.

Exit: send LF or <QUIT>.



E50…E59

I/O-Test error

E60…E82

RS232-Test error


Device-specific errors:

E121

Measuring point list full

E137

Method memory full

E154

Temperature not reached / HEATER timeout

E163

Gas flow too low

E165

Max. temperature reached

E168

Temperature sensor defect

E169 Flowmeter

defect

E201

Changer function error (div.)

E202

Dosimat function error

E206

Gas flow too high

E207

Error list full

E208 SCAN

timeout

774 Oven Sample Processor, Instrucciones para el uso

102

Advertising