Metrohm 774 Oven Sample Processor Manual del usuario

Página 117

Advertising
background image

5.10 The remote control tree

:

.Lift

Moving the lift $G,

$S 5.11.6.3

.Station Lift

address

1, 2, ∗

│ ├

.Way

Position

work, rinse, shift, special,

│ └

rest, 0...325 mm

.Heater

Temperature control $G

5.11.6.4

.Temp Target

temperature

INIT, off, 50…250

│ ├

.Time Heating

interval

blank, 1...999 min

.Flow

Gas flow control $G

5.11.6.5

.Address

Pump or valve

Pump, Valve

│ ├

.State Status

on, off

.Dos

Initiate dosing functions $G,

$S

5.11.6.6

.Address

Address of dosing unit

1...12

│ ├

.Value

Volume or function

±0.001...1...999.999 ml, fill,

│ └

release, prepar., empty, eject,

adjust, level

.Scan

Scanning the interfaces

$G, $S

5.11.6.7

.Address

Selection of interface

Rm, RS

│ ├

.Pattern

Input signal or data

│ └

for Rm (Remote):

8 x 1, 0 or * (bin)

ready1, end1, Cond ok,

Cond 737, no error

for RS (RS232):

14 ASCII characters

.Ctrl

Interface control $G

5.11.6.8

.Address Interface

selection

Rm, RS

│ ├

.Pattern

Output signal or data

│ └

for Rm (Remote):

14 x 1, 0 or ∗ (bin),

START device1, START dos1

STOP device1, START 737

ENTER, INIT

for RS (RS232):

14 ASCII characters, &M;$G

.Def

Re-definitions $G 5.11.6.9

.Object Item

selection

DOSRATE, FILLRATE,

│ ├

LIFTRATE, SHIFTRATE,

│ │

DRIVE.PORT

│ │

.Adress

Component address

dependent upon item

│ ├

.Value

Value

dependent upon item

.Wait

Waiting time

$G, $S, $H, $C

5.11.6.10

.Time Waiting

time

0...1...9999 s

.End

Changer RACK / RESET

$G

5.11.6.11



774 Oven Sample Processor, Instrucciones para el uso