5 Descripción detallada

5.12.3 Pin Assignment


RS232C Interface

Transmitted Data (TxD).

If no data are transmitted, the line is held in the “ON” condi-

tion. Data will only be sent when CTS and DSR are in the

”ON” condition and DCD is in the “OFF” condition.

Received Data (RxD)

Data will only be received when DCD is “ON”.

Request to Send (RTS)

ON condition: the changer is ready to send data.

Clear to Send (CTS)

ON condition: Remote station is ready to receive data.

Data Set Ready (DSR)

ON condition: The transmission line is connected.

Signal ground (GND)

Data Carrier Detect (DCD)

ON condition: The level of the received signal is within the

tolerance range (remote station is ready to send data).

Data Terminal Ready (DTR)

ON condition: Instrument is ready to receive data.

Pin 2

Transmitted Data

Pin 3

Received Data

Pin 4

Request to Send

Pin 5

Clear to Send

Pin 6

Data Set Ready

Pin 7

Signal Ground

Pin 8

Data Carrier Detect

Pin 20

Data Terminal Ready


Protective earthing
Direct connection from cable plug to the protective ground of the instrument.

Polarity allocation of the signals
-

Data lines (TxD, RxD)

voltage negative (<–3 V): signal state "ON"

voltage positive (>+3 V): signal state "ZERO"

-

control or message lines (CTS, DSR, DCD, RTS, DTR)

voltage negative (<–3 V): OFF state

voltage positive (>+3 V): ON state

In the transitional range from +3 V to –3 V the signal state is undefined.

Driver 14C88

according to EIA RS232C specification

Receiver 14C89 " "