5.9 Operation via RS232 Interface

Detailed Status Conditions

Status conditions of the global $R:

$R.Mode

Basic state: ready to start automatic processing

$R.Assembly

An assembly step has been executed.

Status conditions of the global $G:

$G.Mode.Start.

Instrument at the beginning of processing

$G.Mode.Start.01.WAIT

Instrument processing the start sequence, displays line number and
current command

$G.Mode.Sample.01.WAIT

Instrument processing the sample sequence, displays line number and
current command

$G.Mode.Final.01.WAIT

Instrument processing the final sequence, displays line number and
current command

$G.Mode.

Instrument processing a manual command

$G.Assembly.

Instrument processing an assembly-command

Status conditions of the global $H:

$H.Mode

The status conditions of the global $H are identical with the ones of the
global $G.

5.9.5 Error Messages, Errors


Fatal Instrument Errors:

E1

Incorrect program check sum

E2

RAM read/write error

E3

RAM lost data

E4

Timer interrupt for multi-tasking missing

E5

RS232 module test error

E6

RS232 read/write error

E7

Display read/write error

E12 EBUS

error

E18

Low battery

E19

RAM test error


Program-specific Messages:

E26

Manual stop

E28

Wrong object call

E29

Wrong value or no value allowed

E30 Wrong

trigger

E31

Comman currently not allowed


RS Receive Errors:

E36 Parity

error

Exit: <QUIT> and set the same parity for both instruments.

E37

Stop bit error
Exit: <QUIT>and set the same stop bit for both instruments.

E38

Overrun error. At least 1 character could not be read.
Exit: <QUIT>.

E39

Internal receive buffer full (>82 characters).