5.11 Description of the remote control commands

5.11.6 &Assembly …

5.11.6.1

Assembly.Sample $G

Assembly.Sample.Func

=, +, –

Assembly.Sample.Value 0...1...999
Defines the (first) sample beaker (rack position) to be processed. Modifica-
tion of the sample variable.
.Func Function
.Value

Value (absolute or relative)

&Assembly.Sample;$G triggers this function.

5.11.6.2

Assembly.Move $G,

$S

Assembly.Move.Target 1,

2

Assembly.Move.Position

sample, spec.1…8, 0...999

Positioning a beaker at the specified tower.
.Target

Target or tower

.Position

Rack position or identification of a specific vial

&Assembly.Move;$G triggers this function.

5.11.6.3

Assembly.Lift $G,

$S

Assembly.Lift.Station

1, 2, ∗

Assembly.Lift.Way

rest, work, rinse, shift, special, 0...325 mm

Move lift.
.Station

Selection of lift station (

= both lifts)

.Way

absolute lift position

&Assembly.Lift;$G triggers this function.

5.11.6.4

Assembly.Heater $G

Assembly.Heater.Temp

Init, off, 50…250

Assembly.Heater.Time

blank, 1...999

Temperature control.
.Temp

Target temperature in °C

.Time

Heating interval in minutes

&Assembly.Heater;$G triggers this function.

5.11.6.5

Assembly.Flow $G

Assembly.Flow.Address

Pump, Valve

Assembly.Flow.State

on, off

Pump or valve control.
.Address

Selection of pump or valve

.Value State

&Assembly.Flow;$G triggers this function.

5.11.6.6

Assembly.Dos $G,

$S

Assembly.Dos.Address

1...12

Assembly.Dos.Value

±0.01…999.999 ml, fill, release,

prepar., empty , eject, adjust, level

Control of dosing units.
.Address

Selection of dosing unit

.Value

Volume or function

&Assembly.Dos;$G triggers this function.

774 Oven Sample Processor, Instrucciones para el uso

125