7 &diagnosis – Metrohm 774 Oven Sample Processor Manual del usuario

Página 133

Advertising
background image

5.11 Description of the remote control commands

5.11.7 &Diagnosis …

5.11.7.1

Diagnose.Init $G

Diagnose.Init.Select

param, config,

setup, assembly, all

RAM initialization. Sets all default values for the selected sub-branch. See
also 5.11.4.11.
&Diagnose.Init;$G triggers the initialization..

5.11.7.2

Diagnose.RamTest $G

Diagnose.LcdTest

$G, $S, $H

Diagnose.ContrastTest $G,

$S

Diagnose.KeyTest $G,

$S

Diagnose.IoTest $G,

$S

Diagnose.RsTest $G,

$S

Diagnose.EbusTest $G,

$S

Diagnose.BeeperTest $G,

$S

Diagnose.RackcodeTest $G,

$S

Diagnose.FunctionTest

Diagnostic functions. These functions can be triggered with $G. They may
be stopped with $S.

.RamTest

Working memory test

.LcdTest Display

test

.ContrastTest

Display contrast test

.KeyTest

Displays key code and function of each key

.IoTest

Remote interface test

.RsTest

(This function is not applicable via RS232 interface)

.EbusTest

External Bus interface test

.BeeperTest

Test of the acoustic signal (only with
Config.Beeper"on", see 4.10.2.14)

.RackcodeTest Test of rack code recognition
.FunctionTest

Metrohm internal test

5.11.7.3

Diagnose.SimulateKey

0, 1…6, 8…31

Key simulation (Key codes see 5.11.4.2).

5.11.7.4

Diagnose.InstrNo
Instrument number. This entry is not accessible via RS232 interface.

5.11.7.5

Diagnose.OvenTest $G,

$S

Diagnose.Adjust $G,

$S

Diagnose.MeasTest $G,

$S

.OvenTest Heater

test.

.Adjust

Adjustment of temperature and gas flow measurement.

.MeasTest

Test of temperature measurement.

5.11.7.6

Diagnose.PowerOn $G
Simulation of 'Power on'.


774 Oven Sample Processor, Instrucciones para el uso

127

Advertising