Quick Installation Guide - FM-approved MJK Ultrasonic Sensors

Class I, II and III, Div. 1 & 2,
A, B, C, D, E, F, G
hazardous location

To MJK transmitter

Max. 18 inches

User supplied conduits
and materials, see notes.

Integral sensor cable,
see notes.

Required for Class I, II and III,
Div. 1 & 2 locations sealing fittings,
sealing cement and fiber kit.
See notes

Class I, II and III, Div. 1 & 2,
A, B, C, D, E, F, G
hazardous location

Optional MJK
extension cable
p/n 690010

User supplied conduit and
connection box, see notes.

Required for Class I, II and III,
Div. 1 & 2 locations sealing fittings,
sealing cement and fiber kit.
See notes

To MJK ultrasonic
sensor

Notes for customer supplied materials and services

Cables, fittings and conduits must be installed by the customer in accordance with NEC 501-4, 502-4 or 503-3.
In Class I, II and III, Div. 1 & 2, A, B, C, D, E, F, G hazardous (classified) locations all seal fittigs, sealing com-
pounds, connections boxes, conduits, fittings, etc. must be certified and approved for use in the above men-
tioned locations.
Minimum thread engagement between all threaded joints must be a maximum of 5 full threads.
Interconnecting cable conduits and fittings must be grounded to a proper electrical ground. Bonding between all
conduit connections must be provided and installed by the customer as part of installation.
The sensors are provided with an integral cable. An extension cable must MJK cable, part no. 690010.
The cable must be run in accordance with NEC (ANSI/NFPA 70), CEC pt. I and/or applicable local code require-
ments.
This installation guide is under MJK control, and modifications are not allowed without consent from the certifying
authority.

1.
2.

3.
4.

5.
6.

7.