Apéndice E. Test Certificate
Part no.:
Serial no.:
Calibration code:
Size:
Pressure rating:
Protection:
Lining:
Electrodes:
Media temp.:
207222
au2921
23734
IP67
Hard rubber
AISI 316 TI
-10 to 80°C
31 January 2007
DN50
PN16
Test conditions:
The meter tests were carried by a volume method with flying start.
Test facility:
Helios AG, SRN electro-mechanical weighing machines of capacities
4800 kgs production series numbers 9229012, 47583.
Last calibration date 27th December 2006, calibration sheets No. 305-KL-V347-06.
Test results:
Deviation 0,09 % at a velocity of 0,5 m/s
Deviation 0,01 % at a velocity of 1,5 m/s
Deviation 0,00 % at a velocity of 6,5 m/s