background image

4.4. Pin assignment of the RS232 / I/O socket

765 Dosimat

48

4.4.2 RS232 interface, 25 and 8 pin sockets


Transmitted Data (TxD).
If no data are transmitted, the line is held in the “ON” condi-
tion. Data will only be sent when CTS is in the ”ON” condi-
tion.



Received Data (RxD)




Request to Send (RTS)
ON condition: Titrino is ready to send data.



Clear to Send (CTS)
ON condition: Remote station is ready to receive data.



Signal ground (GND)




Data Terminal Ready (DTR)
ON condition: Instrument is ready to receive data.


Data Set Ready (DSR)
ON condition: Communication channel is connected.



Received Line Signal Detector (RLSD)
ON condition: The connected device is ready to send data.

Protective earthing
Direct connection from cable plug to the protective ground of the in-
strument.

Polarity allocation of the signals
- Data lines (TxD, RxD)

voltage negative (<-3 V): signal state ”ON”

voltage positive (>+3 V): signal state ”ZERO”

- Control or message lines (CTS, RTS, DTR, DSR, RLSD)

voltage negative (<-3 V): OFF state

voltage positive (>+3 V): ON state

In the transitional range from +3 V to -3 V the signal state is unde-
fined.