Metrohm 765 Dosimat Manual del usuario

Página 43

Advertising
background image

4.2 Control commands

765 Dosimat

39

4.2.22 PSM VALUE

Parameter 'smpl', live

Value = -1E33 ... -1E-37 , 0 , 1E-37 ... 1E33

Setting sample size.

In mode DOS only, in other modes, the command is not accepted and bit 0 of information

byte 2 is set to 1.

4.2.23 UNI X Unit,

live

X = 0 ... 9, J, K

Selecting the unit:

X = 0, %

X = 1, g

X = 2, mg

X = 3, g/l

X = 4, mg/l

X = 5, mol

X = 6, mol/l

X = 7, ml

X = 8, l

X = 9, /pc (per piece)

X = J, no unit

X = K, ppm

In mode DOS only, in other modes, the command is not accepted and bit 0 of information

byte 2 is set to 1.

4.2.24 VUP VALUE

Rate up, live

Value = 0.001 ... 150.0

Setting rate up in ml/min.

If rate up is set to 'analogue', it is changed automatically to rate up digital.

Input range depends on the cylinder volume of the exchange unit:

Cylinder

ml

rate

min

ml/min

rate

max

ml/min

1 0.001 3

5 0.005 15

10 0.010 30

20 0.020 60

50 0.050 150

If a value is entered which is too high or too low, resp., the Dosimat corrects the value

automatically to rate

max

or rate

min.

resp. and sets bit 1 of information byte 2 to 1 (see

4.2.7). To set rate

max ,

it is therefore possible to always enter 150 ml/min.

4.2.25 VDW VALUE

Rate down, live

Value = 0.001 ... 150.0

Setting rate down in ml/min.

If rate down is set to 'analogue', it is changed automatically to rate down digital.

Input range depends on the cylinder volume of the exchange unit: