Metrohm 765 Dosimat Manual del usuario

Página 41

Advertising
background image

4.2 Control commands

765 Dosimat

37

Information byte 2:

Bit Function

0

1= Wrong command code

1

1 = Parameter corrected to its limit value

2

1 = Repeat command in the READY state

3

1 = Cylinder empty

4

1 = Remote control on

5

1 = Data transfer on (send RS232 on)

6 Reserve


4.2.8 DOS

Mode DOS, not live

Selection of standard mode DOS. The standard parameters (depending on the exchange

unit) are loaded into the working memory. If the exchange unit is not filled, filling is

executed.

4.2.9 DIS R

Mode DIS R, not live

Selection of standard mode DIS R. The standard parameters (depending on the

exchange unit) are loaded into the working memory. If the exchange unit is not filled,

filling is executed.

4.2.10 DIS C

Mode DIS C, not live

Selection of standard mode DIS C. The standard parameters (depending on the

exchange unit) are loaded into the working memory. If the exchange unit is not filled,

filling is executed.

4.2.11 PIP

Mode PIP, not live

Selection of standard mode PIP. The standard parameters (depending on the exchange

unit) are loaded into the working memory. If the exchange unit is not filled, filling is

executed.

4.2.12 DIL

Mode DIL, not live

Selection of standard mode DIL. The standard parameters (depending on the exchange

unit) are loaded into the working memory. If the exchange unit is not filled, filling is

executed.

4.2.13 MDO

Mode DOS with previous parameters, not live

Selection of mode DOS without changing the actual parameters in the working memory.

No filling of the exchange unit.

4.2.14 MDR

Mode DIS R with previous parameters, not live

Selection of mode DIS R without changing the actual parameters in the working memory.

No filling of the exchange unit.