Sample, Apéndice e. test certificate, Apéndice d. dibujo de recorte del panel frontal – Xylem Magflux Electromagnetic Flow Meter Manual del usuario

Página 119: Cut-out ar ea frontpanel outline up

Advertising
background image

119

MagFlux Medidor electromagnético de caudal

FW: 842018/842009

Apéndice E. Test Certificate

Part no.:
Serial no.:
Calibration code:

Size:

Pressure rating:

Protection:

Lining:
Electrodes:
Media temp.:

207222

au2921

23734

IP67

Hard rubber

AISI 316 TI

-10 to 80°C

31 January 2007

DN50

PN16

Test conditions:

The meter tests were carried by a volume method with flying start.

Test facility:

Helios AG, SRN electro-mechanical weighing machines of capacities
4800 kgs production series numbers 9229012, 47583.
Last calibration date 27th December 2006, calibration sheets No. 305-KL-V347-06.

Test results:

Deviation 0,09 % at a velocity of 0,5 m/s
Deviation 0,01 % at a velocity of 1,5 m/s
Deviation 0,00 % at a velocity of 6,5 m/s

Sample

Apéndice D. Dibujo de recorte del panel frontal

La línea de puntos indica el borde del panel frontal y mide 155 x 145 mm.

Cut-out ar

ea

Frontpanel outline

Up