GUÍA DEL USUARIO _______________________________________________________________
122 _________________________________________________________________ M210855ES-D
Ingresar datos de ajuste MPCP1, seis puntos:
>MPCP1
1. Reading ? 800.21
1. Reference ? 800.10
2. Reading ? 850.22
2. Reference ? 849.2
3. Reading ? 901.02
3. Reference ? 899.99
4. Reading ? 949.98
4. Reference ? 950.01
5. Reading ? 1000.11
5. Reference ? 1000.10
6. Reading ? 1050.12
6. Reference ? 1050.12
7. Reading ?
(Dejar vacío para finalizar la entrada de datos).
Multipoint adjustment: OFF
Reading Reference Correction
800.21 hPa 800.10 hPa -0.110 hPa
850.22 hPa 849.20 hPa -1.020 hPa
901.02 hPa 899.99 hPa -1.030 hPa
949.98 hPa 950.01 hPa 0.030 hPa
1000.11 hPa 1000.10 hPa -0.010 hPa
1050.12 hPa 1050.12 hPa 0.000 hPa
0.00 hPa 0.00 hPa 0.000 hPa
0.00 hPa 0.00 hPa 0.000 hPa
>
Activar el ajuste MPCP1:
>MPCP1 ON
Multipoint adjustment: ON
>