83

11 Anexo

VEGAFLEX 86 • Protocolo Modbus y Levelmaster

42288-ES-130416

Set Baud Rate

Request:

Parámetro

Length

Code/Data

Set Baud Rate

8 (12) characters ASCII

UuuBbbbb[b][pds]
Bbbbb[b] = 1200, 9600 (default)
pds = parity, data length, stop bit

(optional)
parity: none = 81, even = 71 (de-

fault), odd = 71

Response:

Parámetro

Length

Code/Data

Set Baud Rate

11 characters ASCII

Ejemplo: U01B9600E71
Cambiar equipo en la dirección 1 a la tasa de baudios 9600, paridad par, 7 bits de datos, 1 bit de

parada

Set Receive to Transmit Delay

Request:

Parámetro

Length

Code/Data

Set Receive to Transmit Delay

7 characters ASCII

UuuRmmm
mmm = milliseconds (50 up to 250),

default = 127 ms

Response:

Parámetro

Length

Code/Data

Set Receive to Transmit Delay

6 characters ASCII

UuuROK

Report Number of Floats

Request:

Parámetro

Length

Code/Data

Set Receive to Transmit Delay

4 characters ASCII

UuuF

Response:

Parámetro

Length

Code/Data

Set Receive to Transmit Delay

5 characters ASCII

UuuFn
n = number of measurement values

(0, 1 or 2)

Report Receive to Transmit Delay

Request:

Parámetro

Length

Code/Data

Report Receive to Transmit Delay

4 characters ASCII

UuuR