Set receive to transmit delay, Report number of floats, Report receive to transmit delay – VEGA VEGAPULS 63 (≥ 2.0.0 - ≥ 4.0.0) Modbus and Levelmaster protocol Manual del usuario
Página 76: Código de error
76
12 Anexo
VEGAPULS 63 • Protocolo Modbus y Levelmaster
41364-ES-130411
Cambiar equipo en la dirección 1 a la tasa de baudios 9600, paridad par, 7 bits de datos, 1 bit de
parada
Set Receive to Transmit Delay
Request:
Parámetro
Length
Code/Data
Set Receive to Transmit Delay
7 characters ASCII
UuuRmmm
mmm = milliseconds (50 up to 250),
default = 127 ms
Response:
Parámetro
Length
Code/Data
Set Receive to Transmit Delay
6 characters ASCII
UuuROK
Report Number of Floats
Request:
Parámetro
Length
Code/Data
Set Receive to Transmit Delay
4 characters ASCII
UuuF
Response:
Parámetro
Length
Code/Data
Set Receive to Transmit Delay
5 characters ASCII
UuuFn
n = number of measurement values
(0, 1 or 2)
Report Receive to Transmit Delay
Request:
Parámetro
Length
Code/Data
Report Receive to Transmit Delay
4 characters ASCII
UuuR
Response:
Parámetro
Length
Code/Data
Report Receive to Transmit Delay
7 characters ASCII
UuuRmmm
mmm = milliseconds (50 up to 250),
default = 127 ms
Código de error
Error Code
Name
EE-Error
Error While Storing Data in EEPROM
FR-Error
Erorr in Frame (too short, too long, wrong data)
LV-Error