VEGA VEGASON 64 Foundation Fieldbus Manual del usuario
Página 57

Secondary_value_1
Calibration
Lowest Point
Calibration
Highest Point
Sensor_value
cal_level_lo
cal_level_hi
Calibration
Lowest Point
Calibration
Highest Point
Sensor reference plane
cal_level_lo [%]
cal_level_hi [%]
Fig. 32: Ajuste VEGASON 64
Lista de parámetros para la revisión de dispositivo 3.0
El listado siguiente contiene los parámetros más importantes y su significado:
l
primary_value
-
This is the process value after adjustment and Linearization with the status of the transducer
block
l
primary_value_unit
-
Unit code of 'Primary_value'
l
secondary_value_1
-
Value after min./max.-adjustment (level + level offset). Selected as input to AIFB by setting
'C
hannel' = 2. Unit derives from 'Secondary_value_1_unit'
l
secondary_value_1_unit
-
Unit code of 'Secondary_value_1'
l
secondary_value_2
-
Sensor value + sensor offset. Selected as input to AIFB by setting 'Channel' = 3. Unit derives
from 'Secondary_value_2_unit'
l
secondary_value_2_unit
-
Unit code of 'Secondary_value_2'
VEGASON 64 • Foundation Fieldbus
57
10 Anexo
28793
-
ES
-090203