Electromagnetic immunity

The CP 150 electrocardiograph is intended for use in the electromagnetic environment

specified below. The customer or user of the CP 150 electrocardiograph should assure

that it is used in such an environment.

Immunity test

IEC 60601 test level Compliance level

Electromagnetic environment -

guidance

Portable and mobile RF communications

equipment should be used no closer to any

part of the CP 150 electrocardiograph,

including cables, than the recommended

separation distance calculated from the

equation applicable to the frequency of the

transmitter.
Recommended separation distance

Conducted RF
IEC 61000-4-6

3 Vrms
150 kHz to 80 MHz

3 Vrms
150 kHz to 80 MHz

d = (1.2)

Radiated RF
IEC 61000-4-3

3 V/m
80 MHz to 2.5 GHz

3 Vrms
80 MHz to 2.5 GHz

d = (1.2)

80 MHz to 800 MHz

d = (2.3)

800 MHz to 2.5 GHz

where P is the maximum output power

rating of the transmitter in watts (W)

according to the transmitter manufacturer

and d is the recommended separation

distance in metres (m).
Field strengths from fixed RF transmitters,

as determined by an electromagnetic site
survey,

a

should be less than the

compliance level in each frequency range.

b

Interference may occur in the vicinity of

equipment marked with the following

symbol:

Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.

Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is

affected by absorption and reflection from structures, objects, and people.

a

Field strengths from fixed transmitters, such as base stations for radio (cellular/

cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast

and TV broadcast cannot be predicted theoretically with accuracy. To assess the

electromagnetic environment due to fixed RF transmitters, an electromagnetic site

survey should be considered. If the measured field strength in the location in which the

CP 150 electrocardiograph is used exceeds the applicable RF compliance level above,

the electrocardiograph should be observed to verify normal operation. If abnormal

performance is observed, additional measures may be necessary, such as reorienting or

relocating the electrocardiograph.

b

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/

m.

Instrucciones de uso

EMC guidance and manufacturer’s declarations 47