Mapping of process value status – VEGA SOLITRAC 31 Foundation Fieldbus Manual del usuario

Página 73

Advertising
background image

73

11 Anexo

SOLITRAC 31 • Foundation Fieldbus

41781-ES-131223

FF descriptor

Rel. In-

dex

Description

Unit

ERROR_TEXT

45

Error text

-

PMT_VOLTAGE_AC-

TUAL

46

Current photomultiplier voltage

-

STANDARDIZATION_

FACTOR

47

Factor for the real value correction

-

SERIALNUMBER

48

Serial number

-

NAMUR_STATE

49

NAMUR state

-

NULL_COUNT_RATE

50

Zero count rate

-

COUNT_RATE_PMT

51

Pulse rate photomultiplier (raw values)

-

ADJ_DENSITY_ABS_

COEFF

52

Process absorption coefficient

-

DEV_SW_VER_ASCII

53

Software version

-

POINT_LEVEL_AD-

JUST_MODE

54

Point level adjustment mode

-

RELAY_VALUE_SEL

55

Relay basic value

-

DIGITAL_IN

56

Frequency of digital input

-

DIGITAL_IN_BOOL

57

State of digital input

-

ANALOG_IN

58

Current on analog input

-

FF_CHANNEL_AVAI-

LABLE

59

Available channels

-

FF_CHANNEL_USED

60

Used channels

-

Mapping of Process Value Status

Hex

Quality

Sub-Status

Condition

0x00

Bad

Non-specific

Unexpected error

0x01

Bad

Configuration Error

It was tried to set a wrong unit with FF interface
On user error codes: 16, 17, 25, 52, 57, 66, 72,

117, 120

0x0C

Bad

Device Failure

Five or more internal communications have failed

while "Process Data Update". Last usable value is

displayed
On user error codes: 8, 34, 35, 36, 37, 38, 73, 80,

86, 121, 122, 141

0x10

Bad

Sensor Failure

On user error codes: 40, 41, 53, 68, 123, 124, 125

0x1C

Bad

Out of Service

Transducer block is in mode "Out of Service"
Channel is not assigned to an AIFB
Channel is not available in running application

0x20

Bad

Transducer in MAN

Transducer block is in mode "Manual"

0x44

Uncertain

Last Usable Value

Three or more internal communications have fai-

led while "Process Data Update". Last usable value

is displayed

Advertising