11 Anexo

POINTRAC 31 • Foundation Fieldbus

41779-ES-131223

FF descriptor

Rel. In-

dex

Description

Unit

PMT_VOLTAGE_CALI-

BRATION

43

Photomultiplier voltage on delivery

-

CORRELATION_CO-

EFF

44

Correlation coefficient for linearizer table

-

ERROR_TEXT

45

Error text

-

PMT_VOLTAGE_AC-

TUAL

46

Current photomultiplier voltage

-

STANDARDIZATION_

FACTOR

47

Factor for the real value correction

-

SERIALNUMBER

48

Serial number

-

NAMUR_STATE

49

NAMUR state

-

NULL_COUNT_RATE

50

Zero count rate

-

COUNT_RATE_PMT

51

Pulse rate photomultiplier (raw values)

-

ADJ_DENSITY_ABS_

COEFF

52

Process absorption coefficient

-

DEV_SW_VER_ASCII

53

Software version

-

POINT_LEVEL_AD-

JUST_MODE

54

Point level adjustment mode

-

RELAY_VALUE_SEL

55

Relay basic value

-

DIGITAL_IN

56

Frequency of digital input

-

DIGITAL_IN_BOOL

57

State of digital input

-

ANALOG_IN

58

Current on analog input

-

FF_CHANNEL_AVAI-

LABLE

59

Available channels

-

FF_CHANNEL_USED

60

Used channels

-

Mapping of Process Value Status

Hex

Quality

Sub-Status

Condition

0x00

Bad

Non-specific

Unexpected error

0x01

Bad

Configuration Error

It was tried to set a wrong unit with FF interface
On user error codes: 16, 17, 25, 52, 57, 66, 72,

117, 120

0x0C

Bad

Device Failure

Five or more internal communications have failed

while "Process Data Update". Last usable value is

displayed
On user error codes: 8, 34, 35, 36, 37, 38, 73, 80,

86, 121, 122, 141

0x10

Bad

Sensor Failure

On user error codes: 40, 41, 53, 68, 123, 124, 125

0x1C

Bad

Out of Service

Transducer block is in mode "Out of Service"
Channel is not assigned to an AIFB
Channel is not available in running application